Does the intraoperative 3D-flat panel control of the planned implant position lead to an optimization and increased in safety in the anatomically demanding region C1/2?

Abstract

Background

The aim of this study was to evaluate the applicability and advantages of intraoperative imaging using a 3D flat panel in the treatment of C1/2 instabilities.

Materials

Prospective single-centered study including surgeries at the upper cervical spine between 06/2016 and 12/2018. Intraoperatively thin K-wires were placed under 2D fluoroscopic control. Then an intraoperative 3D-scan was carried out. The image quality was assessed based on a numeric analogue scale (NAS) from 0 to 10 (0 = worst quality, 10 = perfect quality) and the time for the 3D-scan was measured. Additionally, the wire positions were evaluated regarding malpositions.

Results

A total of 58 patients were included (33f, 25 m, average age 75.2 years, r.:18–95) with pathologies of C2: 45 type II fractures according to Anderson/D'Alonzo with or without arthrosis of C1/2, 2 Unhappy triad of C1/2 (Odontoid fracture Type II, anterior or posterior C1 arch-fracture, Arthrosis C1/2) 4 pathological fractures, 3 pseudarthroses, 3 instabilities of C1/2 because of rheumatoid arthritis, 1 C2 arch fracture). 36 patients were treated from anterior [29 AOTAF (combined anterior odontoid and transarticular C1/2 screw fixation), 6 lag screws, 1 cement augmented lag screw] and 22 patients from posterior (regarding to Goel/Harms). The median image quality was 8.2 (r.: 6–10). In 41 patients (70.7%) the image quality was 8 or higher and in none of the patients below 6. All of those 17 patients the image quality below 8 (NAS 7 = 16; 27.6%, NAS 6 = 1, 1.7%), had dental implants. A total of 148 wires were analyzed. 133 (89.9%) showed a correct positioning. In the other 15 (10.1%) cases a repositioning had to be done (n = 8; 5.4%) or it had to be drawn back (n = 7; 4.7%). A repositioning was possible in all cases. The implementation of an intraoperative 3D-Scan took an average of 267 s (r.: 232-310 s). No technical problems occurred.

Conclusion

Intraoperative 3D imaging in the upper cervical spine is fast and easy to perform with sufficient image quality in all patients. Potential malposition of the primary screw canal can be detected by initial wire positioning before the Scan. The intraoperative correction was possible in all patients.

Trial registration German Trials Register (Registered 10 August 2021, DRKS00026644—Trial registration: German Trials Register (Registered 10 August 2021, DRKS00026644—https://www.drks.de/drks_web/navigate.do?navigationId=trial.HTML&TRIAL_ID=DRKS00026644)

Peer Review reports

Background

The operative therapy of fractures or other instabilities of the upper cervical spine remains a challenging procedure. Minimal implant-misplacement can cause neurovascular complications or material-loosening as a consequence of multiple drilling procedures [1]. Several procedures for the treatment of upper cervical instabilities have been described such as anterior approaches using combined anterior screw osteosynthesis and anterior transarticular C1/2 screw fixation (AOTAF) as well as posterior procedures according to the technique described by Goel/Harms or transarticular C1/2-stabilization regarding to Magerl [2,3,4,5,6,7,8,9].

All methods have in common that screw placement has to be performed in a technical demanding region with small corridor of safety. Traditionally, screw placement is done using intraoperative anatomic landmarks and fluoroscopic control. Intraoperative fluoroscopy is carried out two-dimensionally, i.e. in two planes. However, newer and more innovative C-arms enable the possibility of generation of intraoperative 3D images. The advantages of this technique in spine surgery have been pointed out in several studies dealing with navigated procedures or intraoperative 3D-control alone [10,11,12,13]. In this study we aimed to check the envisaged implant position by controlling the drill canal before the screw was actually implanted. Furthermore, the aim of this study was to evaluate the applicability and advantages of intraoperative imaging using a 3D flat panel in the treatment of C1/2 instabilities. We hypothesized that the technique would be easy and provide sufficient image quality improving patient safety.

Methods

Patient collective

It is a prospective single center study at a level 1 spine center. This study was approved by the local institutional ethics committee of the medical faculty Leipzig (reference 042/21-ek).

The inclusion criteria were patients who were at least 18 years old with instabilities of the C1/2 region caused by fractures, pseudarthroses, malignancies or rheumatoid arthroses. According to our treatment protocol odontoid fractures type II according to Anderson/D'Alonzo with a typical fracture line (Eysel/Roosen Type A and B) were treated from anterior with one or two lag screw(s). Where there were combined fractures of the odontoid process with or without an anterior or posterior fracture of the C1-arch and an atlanto-odontoid osteoarthritis (AO), AOTAF was performed with one or two odontoid lag screws in addition. In case of a high riding vertebral artery and odontoid fractures with a fracture orientation from anteroinferior to posterosuperior as type C based on the classification of Eysel and Roosen [8], posterior stabilization C1/2 regarding to Goel/Harms [5, 7] was performed. This technique was also used in cases of pseudarthroses, pathological fractures or rheumatoid instabilities of C1/2.

Excluded were patients aged younger than 18 years as well as patients where multilevel stabilizations were needed.

Operative treatment

Anterior procedures [lag screws, combined anterior odontoid and transarticular C1/2 screw fixation (AOTAF)]

The operations and surgical procedures were performed in supine position on a carbon table. An endotracheal anesthesia was done. Initially, in case of fractures closed reduction by traction and reclination was performed under fluoroscopic control if necessary. The patient’s head was placed in an adjusted padded carbon occipital ring. A right-sided anteromedial approach of 5–6 cm length was used in all cases. The further approach and the screw fixation of the odontoid process were done analogous to Etter et al. [9]. Due to lack of space, only one screw was placed into the odontoid process in some of the patients. In case of AOTAF bilateral atlantoaxial lateral mass screws were placed. The entrance points of those screws were right lateral to the odontoid screws medial of the sulci described by Lu et al. [14] and Reindl et al. [2] in contrast to the technique described by Polli et al. [3]. Initially, thin 2 mm Kirschner wires (K-wires) were used. Under fluoroscopic control the entrance point was evaluated about 3 mm lateral to the odontoid midline. The direction was tilted 30°–40° dorsally in the sagittal plane and 20°–25° laterally in the frontal plane. For this purpose the vertebral artery had to be checked preoperatively. Hereafter the 2 mm K-wires were exchanged against thinner 1.2 mm K-wires for better image quality. 2D-fluoroscopic images in a.p. and lateral view were taken and evaluated. Then, the 3D-Scan was performed in an apnoe phase. On the one hand the image quality was evaluated like described below, on the other hand the accuracy of the placed K-wires were checked regarding misplacements. If the position was correct, the cannulated screws (3, 5 mm) were implanted; in case of misplacements a revision of the position was performed.

Posterior stabilization C1/2 regarding Goel/Harms

Following sterile draping, a midline incision was performed from the occiput to the tip of the processus spinosus of C3.

For the insertion of the C1 lateral mass screws, the anatomical landmarks are on the same vertical line as the C2 pedicle lines just below the posterior lamina of C1 and above the C1/2 joint. The lateral mass of C1 is identified above the C1/2 joint. With the 2.0-mm drill guide with oscillating attachment and protective sleeve, the correct direction was identified. The drill was oriented parallel to the posterior C1 arch toward the anterior aspect of C1, in general with 10°–20° ascending direction whereas in the axial plane the drill is oriented 10° toward the midline.

For screw placement in C2, the dissection is carried out over the superior surface of the C2 pars interarticularis. With a small nerve hook or elevator, the medial part of C2 pars interarticularis is identified. The entry point is halfway between the upper and lower articular surface of C2 at the vertical line while bisecting the articular mass. Finally, the drill is oriented 25° upward and 15°–25° medially as checked clinically by palpation of the medial wall of the isthmus of C2 with lateral fluoroscopic control [5, 7].

After the drill canal is prepared, thin 1.2 mm Kirschner wires were put into the canal and the 2D-fluoroscopic images in a.p. and lateral view were taken and evaluated (Image 1). Then, the 3D-scan was performed and the results were evaluated like described below. If the position was correct, the cannulated screws (3, 5 mm) were implanted; in case of misplacements a revision of the position was performed.

Seven experienced surgeons were involved. All of them are certificated spine surgeons.

3D-set up

The 3D image intensifier, Ziehm Vision FD Vario 3D©, (Ziehm Imaging GmbH, Nürnberg, Germany) is positioned on the opposite side, the patient lies on a carbon table; the head is positioned in a carbon head frame.

The scan consisted of 110 single images with a radius of 136° and took 110 s in an apnoe phase of the patient. The obtained 3D-scan with multiplanar reconstructions and an isocentrically acquired image series were controlled and evaluated before implanting the screws.

Image quality

The evaluation of the image quality and diagnostic accuracy was performed using a Dicom viewer (efilm, Merge eMed, Milwaukee, WI) by two independent orthopedic trauma surgeons (UJAS, JSJ; both consultants). The inter-rater reliability was calculated regarding to Cohen’s kappa statistics.

A modified numeric analogue scale (NAS) and point system were used to assess the image quality. The NAS system was used to determine an exact score for an image with 10 for excellent image quality and 0 for artifact or not useable, respectively. Furthermore, the point system regarding to Stübig et al. [15] (Table 1) was used to grade the image quality and clinical applicability.

Table 1 Point System regarding to Stübig et al. [15]

Wire position

The position of all wires was analyzed.

Criteria for a misplaced wire were pedicle penetration of the wires medial, lateral or anterior, furthermore a wire placement not sufficiently positioned in the C1 mass and penetration of the anterior cortical edge of the C1 lateral mass. The criteria were stated before the study.

Based on the experience of the authors, the wires in the dens axis can be evaluated sufficiently under 2D fluoroscopy. This was not always the case in the wires or screws of C1 und C2 and the atlantoaxial screws.

Thus, malposition requiring reposition was defined as a clear pedicle penetration of the wires or wire placement not sufficiently positioned in the C1 mass. In these cases, the wires were removed and re-placement was performed.

In addition, in the authors experience the violence of the atlantooccipital joint by wires and screws cannot sufficiently be seen under fluoroscopic control. Thereby, in cases of wire violation of the atlantooccipital joint, the wires were slightly withdrawn backwards before screw length was measured. Here, no correction in terms of a new placement had to be done.

Indications for additive intraoperative 3D Scan or postoperative CT

In case of malposition another 3D-scan was performed after the screws were implanted. If the wire-position was defined as correct, no additional 3D-scan was performed. Regarding to our institution standard a postoperative CT is only performed in case of neurological complications or problems regarding the respiration. In the patients of the current study no complication occurred.

Statistics

All data underwent statistical analysis using standardized SPSS software 17.0 (SPSS, Inc. Chicago, USA). Statistical analysis was made using descriptive statistics. Fisher’s exact test was used to evaluate any associations between image quality and artifacts. A significance level of 0.05 was used.

Results

Patient collective

In 58 patients (33f, 25 m, average age 75.2 years, r.:18–95) with pathologies of the dens axis (45 type II fractures according to Anderson/D'Alonzo with or without arthrosis of C1/2, 2 unhappy triad of C1/2 (Odontoid fracture Type II, anterior or posterior C1 arch-fracture, Arthrosis C1/2) 4 pathological fractures, 3 pseudarthroses, 3 instabilities of C1/2 because of rheumatoid arthritis, 1 C2 arch fracture) 36 patients were treated from anterior [29 AOTAF (combined anterior odontoid and transarticular C1/2 screw fixation), 6 lag screws, 1 cement augmented lag screw] (Table 2) and 22 patients from posterior (stabilization C1/2 regarding to Goel/Harms [5, 7], Table 3).

Table 2 Patient population with treatment via anterior approach
Table 3 Patient population with treatment via posterior approach (all treated with posterior stabilization C1/2 regarding Goel/Harms [5, 7]

The implementation of the intraoperative 3D-Scan (preparing until image evaluation) took averagely 267 s (r.: 232–310 s). No technical problems occurred.

Image quality

Regarding the NAS in the majority of the cases a very good image quality could be achieved. The median was 8,2 (r.: 6–10). In 41 (70.79%) patients the image quality was 8 or higher, in 17 Patients the image quality was deteriorated (NAS 7 = 16; 27.6%, NAS 6 = 1, 1,7%), all of these patients had dental implants (Table 2). The inter-rater reliability (kappa = 0.79) and the intra-rater (kappa = 0.77) reliability was high. In the point system regarding to Stübig et al. [15] 44.4% of the patients showed excellent quality concerning the subjective image quality, 23.4% good and 32.2% acceptable quality, the median was 1,88 (r.1–4). Concerning the delineation of the cortical bone 15 patients (25.9%) showed an excellent delineation of cortical bone, 26 (44.8%) good and 17 (29.3%) an acceptable delineation, the median was 2,1 (r.:1–3). The articular surfaces could be delineated excellent in 17 patients (29.3%), good in 23 (39.7%), acceptable in 17 patients (29.3%) and barely visible in 1 case (1.7%); the median was 2.1 (r.:1–4). Regarding the artifacts in 10 patients (17.2%) no relevant artifacts were visible. In 30 patients (51.7%) showed few artifacts, in 17 patients (29.3%) moderate artifacts were slightly disturbing. In one case (1.7%) the artifacts were disturbing with a rather limited evaluation. Overall, the median was 2,2 (r.:1–4). Concerning the total clinical assessment 10 patients (17.2%) were evaluated as very good with no open questions, 33 patients (56.9%) were good to evaluate with despite minor quality effects. Finally, 15 patients (25,9%) showed a general possible evaluation. The median was 2,11 (r.: 1–3) Overall, we found a significant reduction of image quality in patients with dental implants. Nevertheless, a reliable assessment of the K-wire placement could be realized in all cases (Fig. 1).

Fig. 1
figure 1

Example of a 2D fluoroscopy with K-Wires in C1

Wire position

Out of a total of 148 intraoperatively controlled K-wires, 133 (89.9%) showed a correct positioning. In the other 15 (10.1%) cases a repositioning had to be done (n = 8; 5.4%) or it had to be drawn back (n = 7; 4.7%). A second 3D scan was performed in those 8 patients in which a malposition of the wires was visible and a new drill channel had to be generated. In the additional 7 patients no additional 3D-scan was performed. As the wires only had to be drawn back. A repositioning was possible in all cases (Fig. 2).

Fig. 2
figure 2

Example for misplacement of K-wire, in this case too medial in C1 left side

Discussion

The operative treatment of pathologies in the upper cervical spine remains a complex procedure. Based on postoperative CTs high frequencies up to 21% of screw breaches through the cortex of the C2 pedicle or C1 lateral are seen, underlining a potential danger to nerves and blood vessels, especially the vertebral arteries and the spinal cord [10, 15]. Jacobs et al. [16] demonstrated that narrow pedicles lead to a higher perforation rate of the pedicle wall. According to their study pedicles of < 6.6 mm turned out to be a risk factor for a perforation of the pedicle wall.

Apart from that, malposition is associated with a potential loss of stability. As only thin corridors exist for the implementation of screws, several drill attempts also can lead to screw dislocations as seen in a recent study to AOTAF [4].

To improve accuracy and safety, computer-aided 3-dimensional (3D) imaging is widely used and has proven to be effective [4, 14, 17,18,19,20,21]. It provides better image quality for the anatomical landmarks intraoperatively, especially the complex atlantoaxial region, compared with conventional fluoroscopy [22]. Moreover, instrumentation in critical anatomical regions can already be controlled intraoperatively. The important role of intraoperative multiplanar imaging after osteosynthesis is supported by the high rate of immediate corrections for 11–39% of other anatomic regions in the literature [18,19,20,21, 23,24,25]. Hence, the additional medical benefit seems undisputed [13, 26].

In this study we proved the applicability and advantages of intraoperative imaging using a 3D flat panel in the treatment of C1/2 pathologies. Instead of controlling the final placement of the screws we scanned the trajectory of the drilling canal in term of the K-wires, which were put into the thin drilling canal. Hence a possible correction after a scan could be easily performed. This might be more difficult when a screw with a thicker diameter has already been implanted. Therfore we tactically moved one step forward regarding the 3D-Scan. In the present study the huge majority, (133, 90%), out of a total of 148 intraoperatively controlled K-wires, were placed correctly and only 10% required repositioning due to not optimal positioning. Repositioning was possible in all cases intraoperatively. Thereby, insufficient screw hold or drastic complications such as neurologic deficits or bleeding could be avoided.

Concerning the image quality our results demonstrated that the average score for the overall image quality was about 8 with the NAS, and mainly very good or good for the point system. The clinical assessment score was good despite minor quality defects. Artifacts mainly were seen when the patients had dental implants. Here, a significant correlation could be found. Nevertheless, misplacements also occurred in cases of good or very good image quality and independently of dental implants, here no significant correlation could be found.

Apart from that we have not seen other factors influencing the image quality such as osteoporosis or obesity [27]. In other parts of the human body like the upper thoracic spine or the pelvis region the image quality can be inferior because of the relatively thick soft-tissue envelope. A scan can also be difficult in obese patients in terms of realizing the rotation of the C-Arm. However, this is no problem in the area of the upper cervical spine.

Compared to other studies we can see an improvement in the image quality over the years. Although it is difficult to find comparable studies dealing with the image quality of 3D-scan of the upper cervical spine, former studies showed worse image qualities. Stübig et al. [15] received an average score for the overall image quality with about 7 with the NAS, and good or acceptable for the point system. The clinical assessment score was good despite minor quality defects.

Another often-discussed issue is the increased time required by the use of intraoperative 3D-image devices. In our study, however, the implementation of the intraoperative 3D-scan (from preparing of the C-Arm until image evaluation) took averagely only 267 s (r.: 232–310 s). We did not face any technical problems. Here we can see a significant improvement over the previous generations of 3D-C-Arms, where technical problems were more common and the duration before the surgeon could start to evaluate the images was longer.

This prolonged operation time also can be seen in combination with navigated procedures. In a former study of the authors the average time for a scan including the preparation was 409 s [1]; in addition we faced technical problems in several cases concerning the combination of 3D-C-Arm and navigation system. Apart from that a prolongation has to be taken into account especially at the beginning of using navigation tools due to the learning curve [28]. Furthermore, accuracy problems can occur in case of instable situations in the upper cervical spine. The reference clamp has to be fixed securely to avoid malposition guided by a false navigation.

Several studies demonstrate the safety of navigated odontoid screw placement using O-arm precursors and suggest possible better outcomes compared with conventional techniques. The majority of these studies have compared fluoroscopy-based techniques with the Iso-C system [26]. For instance, Martirosyan et al. [11] compared biplanar fluoroscopy–guided anterior screw fixation of odontoid fractures in 25 patients with Iso-C-guided anterior screw fixation in 26 patients. Surgical complication rates and clinical outcomes were also similar between the groups.

Martirosyan et al. [11] did not detect significant differences in complication rates between patients in whom the O-arm was used and those in whom it was not used as an adjunct. Similar to other 3D-imaging-assisted studies of anterior odontoid fixation, no screws required immediate replacement.

Regarding to the literature the use of intraoperative 3D-Devices with navigation for stabilizations of the upper cervical is technically feasible and shows an increased accuracy of screw placement [13, 18, 29]. Other current alternatives for safe screw implantation are 3D-printed patient specific templates. Tian et al. [30] compared templates with C-Arm navigation and achieved more accurate C2 pedicle screw placement with the 3D-printed navigation template technology although both 3D-printed navigation template-assisted and C-arm based navigation-assisted C2 pedicle and pars screw placement provided similar safety and clinical efficacy. One disadvantage of the templates is the time needed for the production; therefore urgent cases cannot be treated with these devices. Another problem might be the high costs, which also is one disadvantage of navigation systems. This issue is often used as an argument against 3D-C-Arms, possibly connected with other oblique purchases, for example, a carbon operation table or a carbon head holder, as metal artefacts affect the image quality quite seriously. Hüfner et al. [17] highlighted this topic and concluded that, even though the price of a 3D unit is high, the long-term perioperative costs could be decreased by frequent application of this device and, thereby, higher rates of avoided revisions. However, a substantial deficit could arise if the revision rate is only up to 5%.

Limitations

Our study offers some limitations. Even if a wire is placed correctly, the screw still can be misplaced, especially in osteoporotic bone. Apart from that, the initial placement of the wires or the drilling process can also lead to relevant complications. Here, navigated procedures could bring more safety. Another drawback is the fact that no additive scan was performed after implanting of the screws or a revision regularly. We only perform a second scan in cases of insecure evaluation of the implants, which was not necessary in our study; nevertheless we did not see screw-related complication postoperatively. Apart from that there was no control group, so that we are not in the position to demonstrate the clinical advantage.

Conclusion

Intraoperative 3D imaging in the upper cervical spine is fast and easy to perform with sufficient image quality even in patients with dental implants. A screw canal malposition can be detected by wire positioning in the drill cannel prior 3D scanning. If necessary, intraoperative correction was possible in all patients. Improved drill cannel position could be achieved based on the thin drill canal that was created compared to larger bony defects seen after inserting a definite screw.

Availability of data and materials

The datasets used and/or analysed during the current study are available in the additional material except identifying/confidential patient data.

Abbreviations

AOTAF:

Anterior screw osteosynthesis and anterior transarticular C1/2 screw fixation

AO:

Atlanto-odontoid osteoarthritis

K-Wires:

Kirschner wires

NAS:

Numeric analogue scale

f:

Female

m:

Male

n:

Number

r:

Range

References

  1. Jarvers J-S, Katscher S, Franck A, Glasmacher S, Schmidt C, Blattert T, Josten C. 3D-based navigation in posterior stabilisations of the cervical and thoracic spine: problems and benefits. Results of 451 screws. Eur J Trauma Emerg Surg. 2011;37:109–19. https://doi.org/10.1007/s00068-011-0098-1.

    Article  PubMed  Google Scholar 

  2. Reindl R, Sen M, Aebi M. Anterior instrumentation for traumatic C1–C2 instability. Spine. 2003;28:E329–33. https://doi.org/10.1097/01.BRS.0000083328.27907.3B.

    Article  PubMed  Google Scholar 

  3. Polli FM, Miscusi M, Forcato S, Raco A. Atlantoaxial anterior transarticular screw fixation: a case series and reappraisal of the technique. Spine J. 2015;15:185–93. https://doi.org/10.1016/j.spinee.2014.09.019.

    Article  PubMed  Google Scholar 

  4. Josten C, Jarvers J-S, Glasmacher S, Heyde C-E, Spiegl UJ. Anterior transarticular atlantoaxial screw fixation in combination with dens screw fixation for type II odontoid fractures with associated atlanto-odontoid osteoarthritis. Eur Spine J. 2016;25:2210–7. https://doi.org/10.1007/s00586-016-4493-5.

    Article  PubMed  Google Scholar 

  5. Harms J, Melcher RP. Posterior C1–C2 fusion with polyaxial screw and rod fixation. Spine. 2001;26:2467–71. https://doi.org/10.1097/00007632-200111150-00014.

    Article  CAS  PubMed  Google Scholar 

  6. Guppy KH, Chakrabarti I, Banerjee A. The use of intraoperative navigation for complex upper cervical spine surgery. Neurosurg Focus. 2014;36:E5. https://doi.org/10.3171/2014.1.FOCUS13514.

    Article  PubMed  Google Scholar 

  7. Goel A, Laheri V. Plate and screw fixation for atlanto-axial subluxation. Acta Neurochir. 1994;129:47–53. https://doi.org/10.1007/BF01400872.

    Article  CAS  PubMed  Google Scholar 

  8. Eysel P, Roosen K. Ventrale oder dorsale Spondylodese der Densbasisfraktur—eine neue Klassifikation zur Wahl des chirurgischen Zuganges [Ventral or dorsal spondylodesis in dens basal fracture—a new classification for choice of surgical approach]. Zentralbl Neurochir. 1993;54:159–65.

    CAS  PubMed  Google Scholar 

  9. Etter C, Coscia M, Jaberg H, Aebi M. Direct anterior fixation of dens fractures with a cannulated screw system. Spine. 1991;16:S25-32. https://doi.org/10.1097/00007632-199103001-00006.

    Article  CAS  PubMed  Google Scholar 

  10. Yang YL, Fu BS, Li RW, Smith PN, Mu WD, Li LX, Zhou DS. Anterior single screw fixation of odontoid fracture with intraoperative Iso-C 3-dimensional imaging. Eur Spine J. 2011;20:1899–907. https://doi.org/10.1007/s00586-011-1860-0.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Martirosyan NL, Kalb S, Cavalcanti DD, Lochhead RA, Uschold TD, Loh A, Theodore N. Comparative analysis of isocentric 3-dimensional C-arm fluoroscopy and biplanar fluoroscopy for anterior screw fixation in odontoid fractures. J Spinal Disord Tech. 2013;26:189–93. https://doi.org/10.1097/BSD.0b013e31823f62c7.

    Article  PubMed  Google Scholar 

  12. Verhofste BP, Glotzbecker MP, Hresko MT, MacDougall RD, Birch CM, O’Neill NP, Karlin LI, Emans JB, Proctor MR, Hedequist DJ. Intraoperative use of O-arm in pediatric cervical spine surgery. J Pediatr Orthop. 2020;40(4):e266–71. https://doi.org/10.1097/BPO.0000000000001416.

    Article  PubMed  Google Scholar 

  13. Jacobs C, Roessler PP, Scheidt S, Plöger MM, Jacobs C, Disch AC, Schaser KD, Hartwig T. When does intraoperative 3D-imaging play a role in transpedicular C2 screw placement? Injury. 2017;48(11):2522–8. https://doi.org/10.1016/j.injury.2017.09.008. (Epub 2017 Sep 8).

    Article  PubMed  Google Scholar 

  14. Lu J, Ebraheim NA, Yang H, Heck BE, Yeasting RA. Anatomic considerations of anterior transarticular screw fixation for atlantoaxial instability. Spine. 1998;23:1229–35. https://doi.org/10.1097/00007632-199806010-00011.

    Article  CAS  PubMed  Google Scholar 

  15. Stübig T, Kendoff D, Citak M, Geerling J, Khalafi A, Krettek C, Hüfner T. Comparative study of different intraoperative 3-D image intensifiers in orthopedic trauma care. J Trauma. 2009;66:821–30. https://doi.org/10.1097/TA.0b013e31815edf34.

    Article  PubMed  Google Scholar 

  16. Anderson LD, D’Alonzo RT. Fractures of the odontoid process of the axis. J Bone Joint Surg Am. 1974;56:1663–74.

    Article  CAS  PubMed  Google Scholar 

  17. Hüfner T, Stübig T, Gösling T, Kendoff D, Geerling J, Krettek C. Kosten- und Nutzenanalyse der intraoperativen 3D-Bildgebung. [Cost-benefit analysis of intraoperative 3D imaging]. Unfallchirur. 2007;110:14–21. https://doi.org/10.1007/s00113-006-1202-6.

    Article  Google Scholar 

  18. Tonetti J, Boudissa M, Kerschbaumer G, Seurat O. Role of 3D intraoperative imaging in orthopedic and trauma surgery. Orthop Traumatol Surg Res. 2020;106(1S):S19–25. https://doi.org/10.1016/j.otsr.2019.05.021. (Epub 2019 Nov 13).

    Article  PubMed  Google Scholar 

  19. Naseri Y, Hubbe U, Scholz C, Brönner J, Krüger MT, Klingler JH. Radiation exposure of a mobile 3D C-arm with large flat-panel detector for intraoperative imaging and navigation—an experimental study using an anthropomorphic Alderson phantom. BMC Med Imaging. 2020;20(1):96. https://doi.org/10.1186/s12880-020-00495-y.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Keil H, Luxenhofer M, Vetter SY, Beisemann N, Grützner PA, Franke J. Evaluation of image quality and assessability of a new flat-panel 3D C-arm compared to mobile and fixed computed tomography in posterior spinal fixation. Int J Med Robot. 2021;17(2):e2181. https://doi.org/10.1002/rcs.2181. (Epub 2020 Oct 28).

    Article  PubMed  Google Scholar 

  21. Kirnaz S, Gebhard H, Wong T, Nangunoori R, Schmidt FA, Sato K, Härtl R. Intraoperative image guidance for cervical spine surgery. Ann Transl Med. 2021;9(1):93. https://doi.org/10.21037/atm-20-1101.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Yeom JS, Buchowski JM, Park K-W, Chang B-S, Lee C-K, Riew KD. Lateral fluoroscopic guide to prevent occipitocervical and atlantoaxial joint violation during C1 lateral mass screw placement. Spine J. 2009;9:574–9. https://doi.org/10.1016/j.spinee.2009.02.008.

    Article  PubMed  Google Scholar 

  23. Kerry G, Ruedinger C, Steiner H-H. Intraoperative three-dimensional fluoroscopy after transpedicular positioning of Kirschner-wire versus conventional intraoperative biplanar fluoroscopic control: a retrospective study of 345 patients and 1880 pedicle screws. J Craniovertebr Junction Spine. 2014;5:125–30. https://doi.org/10.4103/0974-8237.142307.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Schouten R, Lee R, Boyd M, Paquette S, Dvorak M, Kwon BK, et al. Intra-operative cone-beam CT (O-arm) and stereotactic navigation in acute spinal trauma surgery. J Clin Neurosci. 2012;19:1137–43. https://doi.org/10.1016/j.jocn.2012.01.020.

    Article  PubMed  Google Scholar 

  25. Atesok K, Finkelstein J, Khoury A, Peyser A, Weil Y, Liebergall M, Mosheiff R. The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Injury. 2007;38:1163–9. https://doi.org/10.1016/j.injury.2007.06.014.

    Article  CAS  PubMed  Google Scholar 

  26. Yang YL, Zhou DS, He JL. Comparison of isocentric C-arm 3-dimensional navigation and conventional fluoroscopy for C1 lateral mass and C2 pedicle screw placement for atlantoaxial instability. J Spinal Disord Tech. 2013;26:127–34. https://doi.org/10.1097/BSD.0b013e31823d36b6.

    Article  PubMed  Google Scholar 

  27. Holly LT, Foley KT. Intraoperative spinal navigation. Spine. 2003;28:S54-61. https://doi.org/10.1097/01.BRS.0000076899.78522.D9.

    Article  PubMed  Google Scholar 

  28. Zou D, Zhang K, Ren Y, Wu Y, Yang Y, Li Y. Three-dimensional image navigation system-assisted anterior cervical screw fixation for treatment of acute odontoid fracture. Int J Clin Exp Med. 2014;7:4332–6.

    PubMed  PubMed Central  Google Scholar 

  29. Czabanka M, Haemmerli J, Hecht N, Foehre B, Arden K, Liebig T, et al. Spinal navigation for posterior instrumentation of C1–2 instability using a mobile intraoperative CT scanner. J Neurosurg Spine. 2017;27:268–75. https://doi.org/10.3171/2017.1.SPINE16859.

    Article  PubMed  Google Scholar 

  30. Tian Y, Zhang J, Liu T, Tang S, Chen H, Ding K, Hao D. A comparative study of c2 pedicle or pars screw placement with assistance from a 3-dimensional (3D)-printed navigation template versus C-arm based navigation. Med Sci Monit. 2019;25:9981–90. https://doi.org/10.12659/MSM.918440.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

JSJ and UAJS defined the study protocol, operated some of the patients, analysed and interpreted the patient data and wrote the manuscript. NvdH, AV, CP, PP and SG operated some of the patients and reorganized the manuscript. CEH operated some of the patients and helped defining the study protocol, analysed the data and reorganized the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to J.-S. Jarvers.

Ethics declarations

Ethics approval and consent to participate

This study has been performed in accordance with the Declaration of Helsinki and was approved by the ethics committee of the medical faculty Leipzig (reference 042/21-ek).

Consent for publication

All individuals have given general written consent to the study and in the use of their data, including imaging, for analysis and publication.

Competing interests

CEH: Royalties of Medacta Int. The other authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Jarvers, JS., Spiegl, U.A.J., Pieroh, P. et al. Does the intraoperative 3D-flat panel control of the planned implant position lead to an optimization and increased in safety in the anatomically demanding region C1/2?. BMC Surg 23, 37 (2023). https://doi.org/10.1186/s12893-023-01934-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12893-023-01934-7

Keywords