Skip to main content

Comparison of freehand technique and a novel laser-guiding navigation system in femoral neck-cannulated screw fixation: a randomized controlled trial

Abstract

Cannulated screw fixation is essential in treating femoral neck fractures, and the widely used freehand technique has several limitations. Therefore, we designed a new laser-positioning and navigation system and compared its efficacy with that of the traditional freehand technique in the cannulated screw fixation of femoral neck fractures. This randomized controlled single-blind trial recruited patients with femoral neck fracture, who were treated using either the newly designed laser-navigation device or the freehand technique. In in-vitro experiments, using the femoral neck model, the laser group was better than the freehand group in terms of operation time (P = 0.0153) and radiation exposure time (P < 0.001). In in-vivo experiments, involving 30 patients (15 in each group), the laser group was better than the freehand group in terms of operation time (P < 0.001), radiation exposure time (P < 0.001), blood loss (P < 0.001) and first success rate (P = 0.03). There was no difference in visual analog scale score, Harris score, and fracture-healing time between the two groups. In conclusion, the novel laser-guiding navigation system resulted in shorter operation time, less radiation exposure, and higher first success rate compared with the freehand technique. Further qualified investigations with a larger number of patients and longer follow-up are required in the future.

Peer Review reports

Introduction

Femoral neck fractures are very common and often result in obvious morbidity and mortality. Almost 2.4 million femoral neck fractures occur annually worldwide [1]. According to the three-point principle, the inverted triangular cannulated screw is the classical method for treating femoral neck fractures [2].

Numerous novel technologies, such as orthopedic surgical robot navigation, computer-aided navigation, and orthopedic minimally invasive intelligent visualization system, have been invented to optimize the process of channel screw placement [3,4,5]. However, including the most widely used freehand method, these technologies cannot satisfy the needs of simplicity, accuracy, and low radiation exposure simultaneously [6].

Based on Desargues’s theorem [7] and the characteristics of X-ray projection [8], and using the “three-line coplanar positioning” method [9], our team integrated the G-arm machine with the self-designed laser navigation device to achieve the effect of visual positioning of the channel-screw path. This system has been proven to be accurate, simple, and to have less radiation exposure through the preliminary in-vitro experiment. This study verifies the safety and effectiveness of using cannulated screw fixation to treat patients with femoral neck fracture, and compares it with freehand methods.

Materials and methods

In this study, a novel laser navigation device of the femoral neck-channel screw was proposed. Experiments were performed in-vitro and in-vivo to verify its efficacy. Figure 1 shows the flow diagram of the study.

Fig. 1
figure 1

Flowchart of the study design, including in-vivo and in-vitro experiments

Description of the novel device

According to Desargues’ theorem of projection geometry and the characteristics of X-ray fluoroscopic imaging, a three-axis co-linear approach (fluoroscopic center axis, target axis and laser positioning axis) is adopted. At the same time, we make use of the visualization characteristics of the laser, when the laser lines representing two planes (Desargues’ plane) cross each other, the confluence is the target channel, to realize the visualization of the three-dimensional positioning of the target channel. The laser-guiding navigation device is composed of two laser pointers (in-line, energy 100 mW, wavelength 520 nm, wavelength 635 nm, Senwei), special laser pointer holder, and locating ring attached to the G-arm’s image intensifier (G-arm Orca, WHALE MEDCINE, Boston, USA). Before operation, the device was adjusted repeatedly to ensure that the two laser lines, X-ray fluoroscopy center line and the target axis, were completely overlapped in horizontal and vertical planes simultaneously (Fig. 2). When this device was first installed in the G-arm’s image intensifier, it required an adjustment time of approximately 10 min. Each G-arm machine required installation and adjustment for the first time and only required approximately 1 min of calibration for subsequent use, and this should be finished before the operation.

Fig. 2
figure 2

Work principle sketch maps of the laser guiding navigation device (A). This meant that the horizontal plane X-ray central fluoroscopy axis of G arm (the green line), vertical plane X-ray central fluoroscopy axis (the red line) and the target screw passage central axis (the blue line) remain coaxial, known as “three-line coaxial” (B, C). The blue tube represented target channel (B)

Laser navigation system in in-vitro experiments

In the sawbone femoral neck model in the blind box, the femoral neck cannulated screw-guide pin was placed using a laser-navigation system and freehand methods, respectively (Fig. 3). Both groups adopted the method of three parallel screws. In the freehand group, the distal guide pin was first placed using G-arm fluoroscopy. After the position was confirmed, the other two guide pins were placed using parallel drill guide. In the laser group, after the laser emitter and positioning ring were calibrated, the distal guide pin was inserted first, under the guidance of laser navigation. After the position was confirmed, the other two guide pins were placed using the parallel drill guide. The operations were all performed by the same doctor, and each group contained 10 models. The operation and radiation time were recorded, and the independent sample t test was used for statistical testing.

Fig. 3
figure 3

The sawbone femoral neck model was firmly fixed in the blind box (A). Ensured that the distal guide pin (Φ = 2 mm) was drilled along the red and green laser lines and confirmed that the Kirschner wire passed through target chanel through fluoroscopy (B). The other two guide pins were also placed by the parallel drill guide

Laser navigation system in in-vivo experiments

Patients

This study enrolled femoral neck-fracture patients who underwent internal fixation with cannulated screws at Beijing Friendship Hospital (Level I trauma center in the region) from June 1st, 2021 to April 1st, 2022. Inclusion criteria: age > 18 years, close femoral neck fracture, and acute fracture (the time interval from injury to operation < 2 weeks). Exclusion criteria: pathological fracture, open fracture, old fracture, multiple fractures of the ipsilateral femur, secondary surgery, and patients who preferred other operations. Informed consent was obtained from all patients or their legal guardians.

This was a randomized controlled single-blind trial. The patients enrolled were randomized into laser group and freehand group (Fig. 1). Randomization seed was specified, and the randomization sequence was generated using the PROC PLAN procedure of SAS 9.4 (SAS Institute, Cary, NC, USA) software with a 1:1 allocation. All operations were performed by two orthopedists who had received the same amount of time training and had completed more than 50 internal femoral neck fracture fixation by themselves in 2 years.

Treatment protocol

All the patients were placed in the supine position on a special traction table using the G-arm machine. In the freehand group, the distal guide pin was first placed under the perspective of the G-arm machine; the other two guide pins were then placed with using the parallel drill guide. After accurate positioning, the screw channels were enlarged using a hollow drill along the guide pin, and three parallel cannulated screws (WASTON, Cannulated Compression Screws, diameter 7.3 mm) of suitable length were inserted for fixation. In the laser group, the positioning ring and laser emitter were calibrated before the operation; the patient was then anesthetized and disinfected, and the distal guide pin was inserted under laser-line guidance. The other two guide wires were also placed using the parallel drill guide. After the position was confirmed using fluoroscopy, three parallel cannulated screws were inserted in the same manner as in the freehand group (Fig. 4).

Fig. 4
figure 4

A The physical photos of the laser guiding navigation device. The horizontal red-light laser pointer and the coronal green-light laser pointer were fixed to the G-arm’s image intensifier by the special round strap. The angle of the laser pointer was determined by the position of the two positioning rings. B X-ray fluoroscopy confirmed that the Kirschner wires were inserted in the ideal position by laser. C X-ray fluoroscopy confirmed that the femoral neck cannulated screws were Inserted

The operation time, radiation-exposure time, first success rate, amount of bleeding, visual analog scale (VAS) score, Harris score, and fracture-healing time were recorded and used for comparison among the two groups. All the patients were monitored at 1, 2, 3, 6, and 9 months postoperatively, or the follow-up ended after fractures healed (Fig. 5).

Fig. 5
figure 5

Preoperative X-ray and three-dimensional CT image of 39 years male with femoral neck fracture (A, B). X-ray image of this patient undertaken femoral neck cannulated screws with the laser guiding navigation device, and the fracture healed after operation 3 months later (C, D)

Statistical analysis

The study was designed to detect, with 90% power and an overall 5% Type I error rate, a 32 s decrement in operation time and 80 s decrement in radiation-exposure time from 172 s to 106 s, as estimated in the freehand group, based on the pilot in-vitro experiment. Thirty subjects were required for the final analysis.

The independent sample t-test, Wilcoxon rank-sum test, Chi-square test, and Fisher’s exact test were used to compare the outcomes between the two groups, where appropriate. All the statistical analyses were performed using the SAS JMP version 16.0 (SAS Institute) software, and statistical significance was defined as P < 0.05.

Results

In-Vitro experiments

The average operative time of the freehand group was 177.5 s, whereas the time was 136.5 s in the laser group. The average radiation-exposure time of the freehand and the laser groups was 105.5 s and 27 s, respectively. All differences were statistically significant (Wilcoxon signed rank test, P < 0.05, Table 1).

Table 1 In-vitro Experiments data. Descriptive statistics for each group presented as Median (Lower Quartile; Upper Quartile), and Wilcoxon signed rank test was applied

In-Vivo experiments

Among the 34 patients admitted to our hospital from June 1st, 2021 to April 1st, 2022, 32 patients who matched the inclusion and exclusion criteria were enrolled (Fig. 1). Sixteen patients were randomized to the laser and freehand groups each. Thirty patients completed the study period. Two patients were lost at postoperative follow-up (one patient in each group was excluded). Table 2 showed the characteristics of patients in the two groups. There was no statistically significant difference in age, sex, causes of injury, and fracture classification (P > 0.05).

Table 2 General characteristics of patients. Categorical data were compared using χ2 tests or Fisher’s exact test. There was no statistically significant difference in two group

The first success rate was 93.3% (14/15) in the laser group and 60% (9/15) in the freehand group; the difference was statistically significant (χ2 = 4.658, P = 0.0309). The operation time of the laser group was 45 (40–45) min, whereas the time was 70 (65–75) min in the freehand group. The radiation-exposure time of the laser and freehand groups was 35 (35–40) s and 110 (95–125) s, respectively. The operation time of the laser group was significantly shorter than that of the freehand group (Z = 4.62991, P < 0.001), and the radiation-exposure time was significantly reduced in the laser group compared to that in the freehand group (Z = 4.68056, P < 0.001). The bleeding amount of the laser and freehand groups were 30 (20–35) mL and 50 (40–50) mL, respectively, and the difference was statistically significant (Z = 3.97943, P < 0.001). There was no statistical difference in VAS score, Harris score, or fracture-healing time between the two groups (Table 3).

Table 3 In-vivo Experiments Section data. Continuous variables with non-normal variables were reported as median (Lower Quartile; Upper Quartile),and categorical data were analyzed using Fisher’s exact test or χ2 tests

Discussion

This study evaluated the efficacy and safety of the laser-guiding navigation system in femoral neck-cannulated screw fixation. Based on Desargues’s projective geometry theorem, the target channel was visualized and located by the intercross laser line using special matching positioning ring. In the in-vitro experiments, we verified the advantages of the device based on the guide pins’ insertion and radiation-exposure time by comparison with the freehand group (Fig. 5).

In the in-vivo study, there was no statistical difference in VAS score, fracture-healing time, and Harris score between the laser and freehand groups. The laser group performed better in terms of operative time, radiation-exposure time, and operative blood loss, which can improve the efficiency of surgery (Fig. 6).

Fig. 6
figure 6

The box plot showed that the laser group was better than the FH group in terms of operation time and radiation exposure time, in-vitro experiments

Through clinical the in-vivo study, the first success rate of the laser group could reach 93.3% vs. 60% in the freehand group, indicating a significant improvement. We further verified the characteristics of its high accuracy (Fig. 7), and provided a novel consideration for solving this technical problem.

Fig. 7
figure 7

The box plot showed that the laser group was better than the FH group in terms of operation time and radiation exposure time and blood loss, in-vivo experiments

Over the last decade, the field of orthopedic navigation has developed rapidly. New technologies have been invented to optimize the femoral neck-cannulated screw-inserted procedure. From the first generation, represented by instrument-positioning device [10,11,12], to the second generation, represented by optical navigation technology of orthopedic surgical robot [4, 13,14,15,16,17,18,19,20], they provide substantial help in orthopedic surgical navigation. However, they have the following drawbacks: cumbersome operation, expensive equipment, inaccuracy, and more radiation exposure (Table 4).

Table 4 Summary of published technologies of femoral neck cannulated screw insert procedure (2000–2021)

Among them, the key characteristics [21] of optical navigation are optical location tracking and image registration, which require technically demanding and dedicated equipment, more time registration, higher cost, and lower penetration [22]. Initially, accord to the Desargues’s theorem [23, 24], we wanted to design the laser-positioning navigation device, which could easily complete the intraoperative positioning accurately and is easy to promote, based on existing equipment [9]. After preoperative installation and debugging, the device can realize real-time calibration by the customized positioning ring, using G-arm. The accuracy and reliability of the device was verified in distal locking of femoral intramedullary nails [6]. Simultaneously, the freehand technique [25] is regarded as the gold standard, when the other methods do not work. Therefore, we selected the freehand technique as the control. Through in-vivo and in-vitro randomized controlled trials, it was proven that the novel device can improve the efficiency of surgery and reduce radiation time compared with freehand means, which provides a new idea for laser positioning and navigation.

Although the findings of this study are persuasive, there were limitations. First, the device could achieve precise location; however how to maintain the stability of guide needle in the process of drilling on non-vertical surface must be further improved. Second, the experience level of the two surgeons was comparable; it was impossible to exclude the bias caused by the difference in technical expertise. Third, although the number of cases had reached the sample size required by statistics, more cases in multiple centers would make the results more convincing.

Conclusion

Compared with the freehand group, the laser-navigation group for the cannulated screw fixation of femoral neck has high surgical efficiency and less radiation-exposure time. This novel device facilitates the accuracy in completing the insertion of femoral neck-cannulated screws and provides a choice for improving the efficiency of surgery.

Availability of data and materials

This study was registered on Chictr.org.cn (23/05/2022, No. ChiCTR2200060236). All data are uploaded to the Clinical Trial Management Public Platform (medresman.org.cn). The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

References

  1. Bhandari M, Swiontkowski M. Management of Acute Hip fracture. N Engl J Med. 2017;377(21):2053–62. https://doi.org/10.1056/nejmcp1611090.

    Article  PubMed  Google Scholar 

  2. Bout CA, Cannegieter DM, Juttmann JW. Percutaneous cannulated screw fixation of femoral neck fractures: the three point principle. Injury. 1997;28(2):135–9. https://doi.org/10.1016/s0020-1383(96)00161-1 .

  3. Muller MC, Belei P, Pennekamp PH, et al. Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study. Int Orthop. 2012;36(7):1463–9. https://doi.org/10.1007/s00264-012-1496-7.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Duan SJ, Liu HS, Wu WC, et al. Robot-assisted percutaneous cannulated screw fixation of femoral Neck fractures: preliminary clinical results. Orthop Surg. 2019;11(1):34–41. https://doi.org/10.1111/os.12430.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Zhao C, Wang Y, Wu X, et al. Design and evaluation of an intelligent reduction robot system for the minimally invasive reduction in pelvic fractures. J Orthop Surg Res. 2022;17(1):205. https://doi.org/10.1186/s13018-022-03089-2.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Gao H, Liu Z, Bai X, et al. Comparison of freehand technique and a novel laser guiding navigation in distal locking of femoral intramedullary nails: a randomized controlled trial. BMC Surg. 2022;22(1):363. https://doi.org/10.21203/rs.3.rs-1690445/v1.

    Article  PubMed  PubMed Central  Google Scholar 

  7. An extension of the theorem of the virial and its application to the kinetic theory of, the constitution of gases. Science. 1883;1(3):65–6. https://doi.org/10.1126/science.ns-1.3.65-c.

  8. Riz APR, Sterchi Y, Schwaninger A. How realistic is threat image projection for X-ray baggage screening? Sens (Basel). 2022;22(6). https://doi.org/10.3390/s22062220.

  9. Gao H, Liu Z, Wang G, et al. A New Accurate, simple and less Radiation exposure device for distal locking of femoral Intramedullary Nails. Int J Gen Med. 2021;14:4145–53. https://doi.org/10.2147/ijgm.s321005.

  10. Browbank I, Bouazza-Marouf K, Schnabler J. Robotic-assisted internal fixation of hip fractures: a fluoroscopy-based intraoperative registration technique. Proc Inst Mech Eng H. 2000;214(2):165–79. https://doi.org/10.1243/0954411001535336.

    Article  CAS  PubMed  Google Scholar 

  11. Muller MC, Belei P, De La Fuente M, et al. Evaluation of a fluoroscopy-based navigation system enabling a virtual radiation-free preview of X-ray images for placement of cannulated hip screws. A cadaver study. Comput Aided Surg. 2011;16(1):22–31. https://doi.org/10.3109/10929088.2010.542694.

    Article  CAS  PubMed  Google Scholar 

  12. Murakami T, Yamazaki K, Ogasa H. ADAPT system is a dramatic advance in computer-assisted Surgery for femoral trochanteric fractures. SICOT J. 2021;7:55. https://doi.org/10.1051/sicotj/2021056.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Wang JQ, Zhao CP, Su YG, et al. Computer-assisted navigation systems for insertion of cannulated screws in femoral neck fractures: a comparison of bi-planar robot navigation with optoelectronic navigation in a synbone hip model trial. Chin Med J (Engl). 2011;124(23):3906–11. PMID: 22442841.

    PubMed  Google Scholar 

  14. He M, Han W, Zhao CP, et al. Evaluation of a Bi-planar Robot Navigation System for Insertion of Cannulated Screws in femoral Neck fractures. Orthop Surg. 2019;11(3):373–9. https://doi.org/10.1111/os.12450.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Schep NW, van Walsum T, De Graaf JS, et al. Validation of fluoroscopy-based navigation in the hip region: what you see is what you get? Comput Aided Surg. 2002;7(5):279–83. https://doi.org/10.3109/10929080209146036.

    Article  CAS  PubMed  Google Scholar 

  16. Kendoff D, Hufner T, Citak M, et al. A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation. Unfallchirurg. 2006;109(10):875–80. https://doi.org/10.3109/10929080601090557.

    Article  CAS  PubMed  Google Scholar 

  17. Moulin B, Tselikas L, De Baere T, et al. CT guidance assisted by electromagnetic navigation system for percutaneous fixation by internal cemented screws (FICS). Eur Radiol. 2020;30(2):943–9. https://doi.org/10.1007/s00330-019-06408-0.

    Article  PubMed  Google Scholar 

  18. Yu T, Yuan BM, Zhang XW, et al. Technology of percutaneous cannulated screw implantation using screw view model of navigation in Garden type I of femoral neck fracture: a case report. Med (Baltim). 2019;98(21):e15591. https://doi.org/10.1097/md.0000000000015591.

    Article  Google Scholar 

  19. Murakami T, Yamazaki K, Ogasa H. ADAPT system is a dramatic advance in computer-assisted surgery for femoral trochanteric fractures. SICOT J. 2021;7:55. https://doi.org/10.1051/sicotj/2021056.

  20. Wang S, Wang B, Lan X, et al. Application of a personalized finite element analysis and 3D-Printed Navigation Template in the treatment of femoral Neck fracture with cannulated screw. Med Biol Eng. 2021;41:934–41. https://doi.org/10.1007/s40846-021-00667-3.

    Article  Google Scholar 

  21. Schreurs R, Baan F, Klop C, et al. Registration-free workflow for electromagnetic and optical navigation in orbital and craniofacial Surgery. Sci Rep. 2021;11(1):18080. https://doi.org/10.21203/rs.3.rs-109225/v1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Chen YX, Zhang K, Hao YN, et al. Research status and application prospects of digital technology in orthopaedics. Orthop Surg. 2012;4(3):131–8. https://doi.org/10.1111/j.1757-7861.2012.00184.x.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Albert A, Adrian, Sandler, Reuben [1968]. An Introduction to Finite Projective Planes. 2015. Dover, ISBN 978-0-486-78994-1.

  24. Bamberg J, Penttila. Tim.Completing Segre’s proof of Wedderburn’s little theorem. Bull Lond Math Soc. 2015;47(3):483–92. https://doi.org/10.1112/blms/bdv021.

    Article  Google Scholar 

  25. Aslan A, Konya MN, Gulcu A, et al. Is electromagnetic guidance system superior to a free-hand technique for distal locking in intramedullary nailing of tibial fractures? A prospective comparative study. Ulus Travma Acil Cerrahi Derg. 2020;26(2):280–6. https://doi.org/10.14744/tjtes.2020.94490.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Additional information

NA.

Funding

This paper was supported by Beijing Municipal Administration of Hospitals Incubating Program Fundings (Code:PX2021004), Capital Medical Development Research Fund(2022-2-2027) and National Natural Science Foundation of China, (82102550).

Author information

Authors and Affiliations

Authors

Contributions

Hua Gao, Baojun Wang contributed to the conception and design of the study. Zhenyu Liu, Hua Gao, and Xiaodong Bai analyzed the data, wrote the manuscript, and drew the figures. Gang Wang, Guoqiang Xu, Ji Ma, Jiatian Wang, Wentao Chen, and Yijun Wang collected and assembled the data. Hua Gao, Zhenyu Liu, and Baojun Wang contributed to the critical revision of the manuscript. All authors contributed to the article and approved the submitted version.

Authors’ information

Dr. Gao Hua works in the field of orthopedic traumatology, with research interests in minimally invasive treatment of extremity fractures and multiple injuries.

Prof. Wang Baojun works in the field of orthopedic traumatology, with research interests in minimally invasive treatment of extremity fractures and multiple injuries.

Corresponding author

Correspondence to Baojun Wang.

Ethics declarations

Ethics approval and consent to participate

The studies involving human participants were reviewed and approved by the ethics committee of Beijing Friendship Hospital(No.2022-P2-199-01) and adhered to the Declaration of Helsinki. All experiments were performed in accordance with relevant guidelines and regulations. The patients/ participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual for the publication of any potentially identifable images or data included in this article.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gao, H., Liu, Z., Bai, X. et al. Comparison of freehand technique and a novel laser-guiding navigation system in femoral neck-cannulated screw fixation: a randomized controlled trial. BMC Surg 23, 319 (2023). https://doi.org/10.1186/s12893-023-02226-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12893-023-02226-w

Keywords