Inflammatory reaction to fish oil coated polypropylene mesh used for laparoscopic incisional hernia repair: a case report

  • Chia Yew Kong1Email author,
  • Lee Lee Lai2,
  • Amanda Yin Yen Khoo3, 4,
  • Nazarina Abdul Rahman5 and
  • Kin Fah Chin6, 7
BMC SurgeryBMC series – open, inclusive and trusted201616:8

https://doi.org/10.1186/s12893-016-0123-8

Received: 23 May 2015

Accepted: 4 February 2016

Published: 11 February 2016

Abstract

Background

Polypropylene meshes are widely used in hernia repairs. Hernia meshes have been developed incorporating coatings of active agents. One commercially available mesh has a fish oil coating which is promoted as having anti-inflammatory properties. We report a case, a symptomatic foreign body granuloma reaction associated with a fish oil coated polypropylene mesh, which required eventual mesh explantation.

Case presentation

A 61-year old lady with previous peptic ulcer disease underwent a laparoscopic intraperitoneal placement of mesh for incisional hernia utilising a fish oil coated polypropylene mesh. The patient presented 3 months after the procedure complaining of dyspepsia and pain at the operative site. There was no discharge. The patient was managed conservatively. She presented 10 months post-operatively with progressively worsening symptoms and a hard palpable mass in the epigastrium. Abdominal laparoscopy revealed dense adhesive disease around the mesh with exudates. Adhesiolysis, mesh explantation and a partial gastrectomy was performed. Histopathological examination revealed a foreign body granuloma formation to the mesh.

Conclusion

In-vivo studies looking at intraperitoneal mesh placement with fish oil coatings including data on surgical outcomes such as fistula and adhesive characteristics are scarce in the literature. Further monitoring and studies are required to investigate the safety and efficacy profile of this mesh type in in-vivo models.

Keywords

Hernia Fish oil coated hernia mesh Hernia repair complications Hernia mesh

Background

Incisional hernias are a common complication after a abdominal surgery, with reported incidence rates ranging from 3 to 20 % [1]. The use of laparoscopy in surgery and laparoscopic intraperitoneal placement of onlay mesh (IPOM) specifically continues to be a matter of debate [1]. Current data indicates that laparoscopic repair of incisional hernias are asssociated with favourable outcomes such as length of stay and infective complications but may be associated with worse cosmetic outcomes [1].

The development of laparoscopic approach presents challenges as direct contact of a mesh with intraperitoneal organs may provoke an inflammatory response which in turn may lead to to adhesion and fistula formation [2]. The C-Qur™ Mesh (Atrium Medical, Hudson, NH, USA), is a polypropylene mesh coated with highly purified pharmaceutical grade fish oil comprising a blend of triglycerides and Omega 3 fatty acids [3]. This mesh belongs to a family of meshes developed with coatings that aim to possess favourable anti-adhesion properties when in contact with intraperitoneal organs.

Case presentation

A 61-year-old Chinese lady presented with a 6-month history of a localised abdominal swelling and distention one and a half years post-midline laparotomy for a bleeding peptic ulcer. Her co-morbidities include a mild hiatus hernia and ischaemic heart disease for which she was on a low-molecular-weight heparin and a platelet inhibitor. She is American Society of Anaesthesiologists (ASA) Grade II and is not on regular non-steroidal, anti-inflammatory medications. On examination, an incisional hernia was found at the previous laparotomy site. It had a smooth surface and a positive cough impulse.

The patient underwent a laparoscopic intraperitoneal placement onlay mesh repair for her hernia. An Optic view port (12 mm) was placed at the left subcostal region as primary port entry with two 5 mm ports at the left flank. Dense abdominal adhesions were found with small bowel adherent to the anterior abdominal wall within the hernia sac. These were localised to the hernia site.

A laparoscopic adhesiolysis was done dissecting the adherent small bowel off the anterior abdominal wall. The hernia defect measured 16 cm by 20 cm. An 20.32 cm by 25.4 cm sized polypropylene mesh coated with omega 3 fatty acids (fish oil) was placed tension-free with no primary suturing and fixed with two rows of ProTack™ 5 mm titanium helical fasteners to restore the integrity of the wall. The hernia sac was not excised. A standard double crown technique with stay sutures was employed. There were no iatrogenic injuries during adhesiolysis. The patient recovered and the immediate post-operative period was uneventful.

During a routine follow-up visit 3 months after the procedure, the patient complained of mild dyspepsia, intermittent abdominal pain. There was no discharge from the incision site. On examination, there was slight tenderness over the incisional hernia site but no obvious recurrence. The patient was reassured and managed conservatively.

The patient’s dyspepsia and abdominal pain got progressively worse and a hard tender mass located in the epigastrium was palpated 10 months post-procedure. Computed tomography imaging of the abdomen showed thickening of the body and pylorus of the stomach and a thin soft tissue plane separating the stomach lesion and the pancreas (Fig. 1a, b).
Fig. 1

(a and b) CT imaging (transverse and sagittal views) showing thickening of the body and pylorus of the stomach (orange arrow) and a thin soft tissue plane separating the stomach lesion and the pancreas. The appearance of tissue extending between the two abdominal wall tacks suggested a recurrence of the incisional hernia but this was found not to be the case intraoperatively. Intraoperatively, the mesh was found to be adherent to a loop of small bowel

Gastroscopy revealed a thickened gastric antrum. There was no gross evidence of acute recurrence of her ulcer disease. Histopathological examination of biopsies taken from the gastric antrum showed chronic inflammatory granulation and fibrosis with no evidence of malignancy.

The patient subsequently underwent a laparoscopy. Dense intrabdominal adhesions were found intra-operatively. There was an adhesion of the small bowel to the fish oil mesh with fine inflammatory exudates. It was converted to an open procedure. Adhesiolysis, explantation of fish oil coated mesh and ProTack™, partial gastrectomy, gastrojejunostomy and jejuno-jejunostomy were performed. There was no evidence for a recurrence of the incisional hernia. There was no evidence of fistula or chronic sepsis. A sutured repair was undertaken. The condition of the patient improved a few days after the surgery.

Two samples were sent for histopathological analysis: the mesh sample and the partial gastrectomy. Macroscopic analysis of the mesh sample revealed the sample to be a torn piece of plastic mesh embedded in fibrous tissue secured by multiple metal coils. Microscopically, sections of the sample showed partly hyalinised fibrofatty tissue exhibiting an intense foreign-body granulomatous inflammatory reaction towards an amorphous translucent foreign material (Fig. 2a, b).
Fig. 2

(a and b) histopathology images showing foreign body granuloma reaction to a pseudomembrane associated with the mesh

The partial gastrectomy specimen had thick haemorrhagic fibrous adhesions. The mucosa showed loss of rugae and there appeared to be superficial ulcerations in the mucosa. Microscopically, the gastric specimens revealed submucosal oedema, inflammed granulation and foci of suppurative inflammatory necrosis with extension of the inflammation across the muscularis into the serosa.

The patient had an unremarkable post-operative recovery and has had no further sequelae in subsequent follow-up visits.

Discussion

The intraperitoneal placement onlay mesh technique involves the placement of mesh within the peritoneal cavity with inevitable contact with adjacent intraperitoneal organs [1]. This contact may provoke an inflammatory reaction between the host tissue and the mesh as a foreign body and may lead to a spectrum of complications from adhesions and fistula formation to bowel obstruction secondary to adhesive disease and organ erosion [2].

These inflammatory reactions are well documented complications of hernia repair with prostheses in both in vitro and in vivo models of hernia repair [2, 47]. Specifically, polypropylene mesh is known to cause a non-infective, host-initiated, foreign body granulomatous inflammatory response which may cause scar thickening [4, 8]. This is hypothesised to result from an interaction between native proteins such as fibrinogen and albumin and the mesh’s polymer, causing a denaturation of these proteins which stimulates the formation of a foreign body granuloma, characterised by the aggregration of macrophages and other inflammatory cells [6]. This process can go on for an extended period of time [5, 6].

Hernia meshes have been developed to possess favourable adhesion and anti-inflammatory properties. Mesh materials for intraperitoneal onlay placement have been developed specifically with coatings which have favourable anti-inflammatory properties required in intraperitoneal onlay meshes [9]. The coatings include polyglactin, hyaluronate and cellulose [9]. The mesh used in this patient is a polypropylene mesh coated with pharmaceutical grade fish oil consisting of a blend of triglycerides and Omega 3 fatty acids [3].

Our patient’s initial recovery from the hernia repair was complicated in two ways, [1] an overt, symptomatic foreign body reaction to the mesh material, forming an inflammatory pseudomembrane and [2] a chronic gastritis. There are a number of explanations which could explain her chronic gastritis. The chronic inflammatory changes in the stomach may be due to her underlying previous peptic ulcer disease. It is however unclear whether this chronic inflammatory gastritis could have alternatively been induced by a direct extension of the inflammatory process from the mesh, although the former is a more likely conclusion. The presence of the abdominal wall lesion and gastric disease gave rise to the patient’s complex presentation, with a decision made to perform a laparoscopy to explore these two issues once the patient’s condition deteriorated.

Our case report highlights the development of a foreign body inflammatory reaction to coated mesh. A few in vitro and in vivo studies involving animal models support the anti-inflammatory characteristics of fish oil and C-Qur specifically. Omega-3 fatty acids have been shown to have anti-inflammatory properties [10]. In a rat model study comparing five coated meshes found that C-Qur provoked the least inflammatory reaction amongst the coated meshes [11]. In a rabbit model, no differences were found in adhesion grade and amount between C-Qur and other meshes but C-Qur and a lower percentage coverage of adhesions [12]. There was a single clinical in vivo human study comparing C-Qur with other commercially available meshes in a prospective study of mesh characteristics in patients undergoing laparoscopy who had a previous mesh implant [9, 13]. C-Qur was found to be associated with the lowest adhesion tenacity, least area covered by adhesions and lowest ratio of adhesiolysis time to adhesion area compared to three other meshes [9, 13]. In a rat model comparing the effect of mesh placement across time however, it was found that while C-Qur significantly reduced adhesion formation at 7 days compared to other meshes, this effect was diminished at 30 days with significant increase in adhesions formation [14]. Therefore there is limited clinical data on fish oil coated meshes in humans in the context of intraperitoneal onlay mesh repairs of incisional hernias. Where evidence exists, these are mostly in animal models with a short-term follow-up. Further, these studies have not been replicated to ensure the validity of these findings. Taken together, however, a review of currently available evidence makes this patient’s response to the mesh unexpected. Within the senior author’s experience (Table 1), this is the only patient in the case series of 11 patients with a foreign body reaction who had a mesh explantation.
Table 1

The single surgeon’s experience with the C-Qur mesh. Patient 3 is the patient described in this case report. NA indicates missing data

No.

Type of surgery

Duration

Complication

Treatment of complication (if any)

Outcome at 6 months if there was complication

1

Laparoscopic incisional

2 h 30 min

Adhesions and seroma

Adhesiolysis

Recovered, well

2

Laparoscopic left inguinal

1 h 50 min

None

  

3

Laparoscopic incisional

NA

Foreign body reaction

Mesh explantation

Recovered, well

4

Laparoscopic incisional

5 h and 35 min

Haematoma

 

Recovered, well

5

Laparoscopic umbilical

2 h 15 min

Haematoma and wound infection

 

Recovered, Well

6

Laparoscopic bilateral inguinal

2 h 30 min

None

  

7

Laparoscopic incisional

2 h 10 min

None

  

8

Laparoscopic bilateral inguinal

2 h and 35 min

None

  

9

Laparoscopic paraumbilical

55 min

None

  

10

Laparoscopic paraumbilical

NA

Wound and mesh infection

Removal of mesh (infected)

Recovered, well

11

Laparoscopic paraumbilical

NA

Seroma

 

Recovered, well

There is a caveat when considering the arguments in this case report. It is likely that a number of factors augment the response of the host to the mesh, which includes host and mesh factors, such as her previous peptic ulcer problem.

Our report calls for further research on the safety and efficacy profile of this mesh type. Ongoing prospective monitoring of the use of this mesh type and the incidence of these complications is warranted.

Conclusion

Polypropylene mesh is known to cause a mild but persistent foreign body reaction but coating with fish oil has been shown to have favourable properties in a few in vivo and in vitro studies, making the findings in this case report unexpected. Data on surgical complications following the usage of fish oil coated mesh for abdominal hernia repair, including infective complications, especially in human studies, however continues to be scarce in the literature. Further studies and monitoring are required to further elucidate the efficacy and safety profile of such meshes.

Consent

Written informed consent was obtained from the patient’s relative/next-of-kin for publication of this Case report and any accompanying images. A copy of the written consent is available for review by the Editor of BMC Surgery.

Declarations

Acknowledgement

We would like to thank the staff of the University of Malaya Medical Centre for help in procuring clinical information and retrieving case files for this case report.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Authors’ Affiliations

(1)
School of Medicine, University of Glasgow
(2)
Department of Nursing Science, University of Malaya
(3)
The University of Queensland Mayne Medical School
(4)
Redland Hospital
(5)
Department of Pathology, University of Malaya, University Malaya Medical Centre
(6)
Department of Surgery, University of Malaya, University Malaya Medical Centre
(7)
Department of Surgery, Tunku Abdul Rahman University

References

  1. Sanders DL, Kingsnorth AN. The modern management of incisional hernias. BMJ. 2012;344, e2843.View ArticlePubMedGoogle Scholar
  2. Leber GE, Garb JL, Alexander AI, Reed WP. Long-term complications associated with prosthetic repair of incisional hernias. Arch Surg. 1998;133:378–82.View ArticlePubMedGoogle Scholar
  3. Atrium Medical Corporation. C-Qur mesh: Instructions for use. 2012. Available at: http://www.atriummed.com/en/biosurgery/Documents/49155_Atrium_CQur_IFU003706.PDF. Accessed date 20 December 2015.
  4. Robinson TN, Clarke JH, Schoen J, Walsh MD. Major mesh-related complications following hernia repair. Surg Endosc. 2005;19(12):1556–60.View ArticlePubMedGoogle Scholar
  5. Brown CN, Finch JG. Which mesh for hernia repair? Ann R Coll Surg Engl. 2010;92:272–8.PubMed CentralView ArticlePubMedGoogle Scholar
  6. Falagas ME, Kasiakuo SK. Mesh-related infections after hernia repair surgery. Clin Microbiol Infect. 2005;11:3–8.View ArticlePubMedGoogle Scholar
  7. Amid PK. Classification of biomaterials and their related complications in abdominal wall hernia surgery. Hernia. 1997;1:15–21.View ArticleGoogle Scholar
  8. Beets GL, van Mameran H, Go PMNYH. Long-term foreign-body reaction to pre-peritoneal polypropylene mesh in the pig. Hernia. 1998;2:153–5.View ArticleGoogle Scholar
  9. Deeken CR, Faucher KM, Matthews BD. A review of the composition, characteristics and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair. Surg Endosc. 2011;26:566–75.View ArticlePubMedGoogle Scholar
  10. Sierra S, Frederico LV, Comalada M, Olivares M, Xaus J. Dietary fish oil n − 3 fatty acids increase regulatory cytokine production and exert anti-inflammatory effects in two murine models of inflammation. Lipids. 2006;41(12):1115–25.View ArticlePubMedGoogle Scholar
  11. Schreinemacher MHF, van Barneveld KWY, Dikmans REG, Gijbels MJJ, Greve JWM, Bouvy ND. Coated meshes for hernia repair provide comparable intraperitoneal adhesion prevention. Surg Endosc. 2013;27:4202–9.View ArticlePubMedGoogle Scholar
  12. Pierce RA, Perrone JM, Nimeri A, Sexton JA, Walcutt J, Frisella MM, et al. 120-Day comparative analysis of adhesion grade and quantity, mesh contraction and tissue response to a novel omega-3 fatty acid bioabsorbable barrier macroporous mesh after intraperitoneal placement. Surg Innov. 2009;16(1):46–54.View ArticlePubMedGoogle Scholar
  13. Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM, et al. Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc. 2010;24:3002–7.View ArticlePubMedGoogle Scholar
  14. Schreinemacher MHF, van Barneveld KWY, Gijbels MJJ, Greve JWM, Beets GI, Bouvy ND. Degradation of mesh coatings and intraperitoneal adhesion formation in an experimental model. Br J Surg. 2009;96:305–13.View ArticlePubMedGoogle Scholar

Copyright

© Kong et al. 2016

Advertisement