Skip to main content

Table 2 Single center, single user bariatrics analysis comparing Vessel Sealer Extend (Erbe VIO da Vinci generator) and SynchroSeal (E-100 da Vinci generator)

From: Novel multifunctional robotically assisted bipolar instrument for simultaneous radiofrequency sealing and transection: preclinical and single-center experience

Procedure and variables

Vessel Sealer Extend

SynchroSeal

p-value

Sleeve gastrectomy

   

 Procedures, n

49

53

 

 Mean activations per procedure, n ± SD

57 ± 25 a

79 ± 26

< 0.001

 Mean accumulated activation time per procedure, s ± SD

249 ± 114 a

119 ± 38

< 0.001

 Mean time tool used per procedure, s ± SD

1030 ± 560

1229 ± 445

0.053

 Mean tool usage as a proportion of time for all tools used per procedure, % ± SD

35 ± 13

43 ± 17

0.007

Gastric bypass (Roux-en-Y)

   

 Procedures, n

33

52

 

 Mean activations per procedure, n ± SD

40 ± 20

74 ± 44

< 0.001

 Mean accumulated activation time per procedure, s ± SD

179 ± 96

108 ± 65

< 0.001

 Mean time tool used per procedure, s ± SD

934 ± 769

1403 ± 809

0.010

 Mean tool usage as a proportion of time for all tools used per procedure, % ± SD

20 ± 13

34 ± 18

< 0.001

  1. SD standard deviation of the mean
  2. a Data missing for 4 cases