Skip to main content

Table 2 Individual measurement parameters for both conventional presurgical CT images and intraoperative O-arm scans

From: The use of an O-arm in endonasal endoscopic operations of the skull base

 

Reference point

Neuronavigation–accuracy error (mm)

O-arm–accuracy error (mm)

Absorbed dose by O-arm (mGy cm)

Total surgical time (min)

Patient 1

Vomer

6.6

0.0

220.03

125

Patient 2

Intrasphenoidal septum

2.8

0.0

220.03

100

 

Intrasphenoidal septum

2.2

0.0

  

Patient 3

Intrasphenoidal septum

1.5

0.0

220.03

105

Patient 4

Intrasphenoidal septum

3.6

0.0

220.03

97

 

Vomer

3.6

0.0

  

Patient 5

Intrasphenoidal septum

0.0

0.0

220.03

66

 

Intrasphenoidal septum

0.0

0.0

  

Patient 6

Intrasphenoidal septum

1.5

0.0

220.03

130

 

Vomer

1.9

0.0