Skip to main content

Table 3 In situ vessel diameters immediately before sealing (Terminal experiment)

From: Safety and effectiveness of a novel generator algorithm for bipolar vessel sealing: a randomised controlled chronic animal study

Vessel type

thermoSEAL mode (TSM)

BiClamp® mode (BCM)

p value

No.

mean ± STD (range) [mm]

No.

mean ± STD (range) [mm]

All vessels

48

4.3 ± 1.4 (2.1–8.1)

49

4.2 ± 1.5 (2.0–11)

0.79

Arteries

31

4.2 ± 0.8 (2.9–6.0)

33

4.1 ± 0.8 (2.5–6.1)

0.97

Veins

17

4.5 ± 2.1 (2.1–8.1)

16

4.4 ± 2.3 (2.0–11)

0.95