Skip to main content

Table 3 Statistical comparison of accuracy between neuronavigation and O-arm

From: The use of an O-arm in endonasal endoscopic operations of the skull base

n = 6

Mean

SD

Minimum

Maximum

Median

p-value

Neuronavigation-deviation (mm)

2.65

2.27

0

6.6

2.1

0.043

O-arm-deviation (mm)

0

0

0

0

0

  1. SD standard deviation