Skip to main content

Table 1 Thermography data

From: Thermal effects of monopolar electrosurgery detected by real-time infrared thermography: an experimental appendectomy study

Group

Low Power

Standard Power

 
 

R1

R3

R6

R8

Mean

(SD)

R2

R4

R5

R7

Mean

(SD)

p

Mean application time, s

2.5

1.46

2.5

1.75

2.05 (0.53)

1.2

1.09

1.52

1.53

1.34 (0.22)

0.112

Min application time, s

0.3

0.1

0.4

0.5

0.33 (0.17)

0.1

0.4

0.4

0.2

0.28 (0.15)

0.773

Max application time, s

11.5

7.9

9.8

9.6

9.7 (1.5)

4.1

3.4

6.8

5.8

5.03 (1.55)

0.03

Application duration > 3 s, %

n/d*

6,3%

25%

6,8%

12.7 (10.7)

3%

1,2%

11,8%

10,3%

6.6

(5.3)

0.595

Cecum T max, °C

58.9

55.3

59.1

48.4

55.4 (4.9)

61.3

57.2

48.8

56.1

55.9 (5.2)

0.885

Mesoappendix T max °C

152.4

146

160.4

157.2

154 (6.3)

160.6

156.4

160.8

160.5

159.6 (2.1)

0.112

T ≥ 42° duration on cecuma, s

42

n/db

13.4

15.5

23.6 (15.9)

24

14

12

13

15.8 (5.6)

0.377

Lateral thermal spreada, mm

8

n/db

4.6

4.48

5.7 (1.9)

5.6

5.6

6

9.3

6.6 (1.8)

0.372

Lateral thermal spread on mesoappendixa, mm

28

24

27.6

20.4

25 (3.6)

33

18.4

20.4

32.4

26.1 (7.7)

0.886

Lateral thermal spread on intestinal loop, mm

1.8

1.6

2.5

3.3

2.3 (0.8)

1.4

1.6

3

6

3 (2.1)

0.886

Complication

no

no

Intestinal wall full thickness necrosis

Intestinal perforation

2/4

no

no

Intestinal perforation

Intestinal perforation

2/4

 
  1. a ≥42 °C
  2. bno data