Skip to main content

Table 1 The mesenteric vascular diameter, thermal damage width and burst pressure for bipolar electrocoagulation under different powers, LigaSure and ultracision

From: Thermal damage width and hemostatic effect of bipolar electrocoagulation, LigaSure, and Ultracision techniques on goat mesenteric vessels and optimal power for bipolar electrocoagulation

Group

Cases

Mesenteric vascular diameter (mm)

Thermal damage width (mm)

Burst pressure (mmHg)

80 W

30

4.194 ± 0.755

7.967 ± 0.350

660.267 ± 38.596

75 W

30

4.230 ± 0.672

7.889 ± 0.359

645.200 ± 36.096

70 W

30

4.231 ± 0.653

6.381 ± 0.406

637.000 ± 42.559

65 W

30

4.228 ± 0.673

6.301 ± 0.378

609.867 ± 30.854

60 W

30

4.311 ± 0.612

5.091 ± 0.332

553.567 ± 37.058

55 W

30

4.188 ± 0.543

5.013 ± 0.320

546.567 ± 34.834

50 W

30

4.204 ± 0.701

6.256 ± 0.339

518.533 ± 25.322

LigaSure

80

4.211 ± 0.678

4.470 ± 0.693

1033.575 ± 82.025

Ultracision

80

4.256 ± 0.605

4.089 ± 0.762

673.538 ± 50.454