Skip to main content

Table 4 Early complications divided in groups according to the implantation technique (Intention To Treat)

From: Totally implantable venous access devices: retrospective analysis of different insertion techniques and predictors of complications in 796 devices implanted in a single institution

Complication

Surgical cutdown (102 pts) SC

Blind percutaneous (48 pts) BP

US guided percutaneous

Fisher exact test p value

Fisher exact test p value

Fisher exact test p value

 

(646 pts) US

SC vs BP

BP vs US

SC vs US

Pneumothorax

0 (0%)

4 (8.3%)

0 (0%)

0.009

<0.001

NA

Arterial puncture

0 (0%)

4 (8.3%)

8 (1.2%)

0.009

0.006

0.390

Arrhythmia

0 (0%)

1 (2.1%)

2 (0.3%)

0.301

0.193

0.179

Technique failure

15 (14.7%)

3 (6.2%)

2 (0.3%)

0.181

0.002

<0.001

Access site change after first attempt

0 (0%)

5 (10.4%)

13 (2%)

0.002

0.005

0.233

Total

15 (14.7%)

17 (35.4%)

25 (3.9%)

0.005

<0.001

<0.001